07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 10-13
10. Operating Instructions – Make Adjustments
10.2.8 Comfort Control
Follow the steps below to adjust the temperature of the heater bag. The range is
from 35°C to 37°C. The default setting is 36°C.
Steps to adjust the temperature Display screen
1. Press
ENTER to access the MAKE ADJUSTMENTS
menu.
2. Press
V until COMFORT CONTROL appears.
3. Press
ENTER.
The temperature digits blink.
(The temperature blinks)
4. Press U or Vto change the temperature.
5. Press
ENTER to save the new temperature.
The blinking stops.
(Blinking stops)
6. Press V to display the next option.
– OR –
7. If you do not want to make any other adjustments,
press
STOP to exit Make Adjustments.
MAKE ADJUSTMENTS
COMFORT CONTROL: 36
COMFORT CONTROL: 36
COMFORT CONTROL: 37
COMFORT CONTROL: 37
LAST MANUAL DRAIN: N