Red Cycler 5C Instrukcje Operacyjne Strona 176

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 175
11. Operating Instructions – Prepare for Therapy
11-2 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
WARNING
Do not use a Patient Line Extension with the Low
Recirculation Set. Using an extension with this set
increases the recirculation volume and reduces therapy
effectiveness.
WARNING
Do not extend the patient line beyond 34 feet (10.4 meters)
for the standard set. Extending the patient line beyond this
length increases the recirculation volume and reduces
therapy effectiveness.
WARNING
Do not use a Patient Line Extension Set if tip protectors are
not in place. If the tip protectors are not secure, possible
contamination of the fluid or fluid pathways can result.
Contamination of any portion of the fluid or fluid path can
result in peritonitis.
WARNING
If the solution is not clear, do not use and discard the bag.
Follow the labeling instructions supplied with the dialysis
solution for storage and preparation. Failure to follow the
solution labeling instructions can lead to insufficient
therapy or adverse clinical reaction.